  1. <?xml version="1.0"?>
  2. <!DOCTYPE task PUBLIC "-//OASIS//DTD DITA Task//EN" "task.dtd">
  3. <task >
  4.   <title>Unloading Bottles Automatically</title>
  5.   <taskbody>
  6.          <context >
  7.                 <p>The
  8.                   <ph
  9.                         conref="../../../../common/col_BMXProductNames.xml#concept_C8D50348E99740B484761C442690972C/prod_NewBacTAlert"><?xm-replace_text Phrase?></ph>can
  10.                   automatically unload all positive and negative bottles from any instrument
  11.                   within a bank when the instrument is configured to automatically unload bottles
  12.                   when testing is complete and a final test result is available.
  13.                 </p>
  14.                 <note>Positive bottles are unloaded before negative bottles.
  15.                 </note>
  16.          </context>
  17.          <steps >
  18.                 <step >
  19.                   <cmd>If the final result is positive, the
  20.                          <b>Unloading Positive Bottles</b> indicator is displayed. The
  21.                          instrument automatically unloads all positive bottles from the cells and
  22.                          transfers the bottles to the Bottle Retrieval Area at the front of the
  23.                          instrument to stop incubation.
  24.                   </cmd>
  25.                 </step>
  26.                 <step >
  27.                   <cmd>Remove the positive bottles from the Bottle Retrieval Area.
  28.                   </cmd>
  29.                   <stepresult>Once the positive bottles are removed from the Bottle
  30.                          Retrieval Area, the positive bottle counts displayed on the Home screen are
  31.                          updated, and the unload events are recorded in the Audit Log.
  32.                   </stepresult>
  33.                 </step>
  34.                 <step >
  35.                   <cmd>If the final result is negative, the
  36.                          <b>Unloading Negative Bottles</b> indicator is displayed and the
  37.                          instrument automatically unloads all negative bottles from the cells and
  38.                          transfers the bottles to the waste container.
  39.                   </cmd>
  40.                   <info>
  41.                          <note>Any alarm or alert associated with an unloaded bottle is
  42.                                 canceled by the instrument, and the event is recorded in the Audit Log.
  43.                          </note>
  44.                   </info>
  45.                   <stepresult>The negative bottle counts displayed on the Home screen are
  46.                          updated, and the unload event is recorded in the Audit Log.
  47.                   </stepresult>
  48.                 </step>
  49.          </steps>
  50.          <postreq >
  51.  
  52.                 <p>Automatic unloading will be stopped and an alarm displays if one or
  53.                   more of the following problems are encountered:
  54.                 </p>
  55.                 <ul >
  56.                   <li >Waste container is full.
  57.                   </li>
  58.                   <li >Waste container is
  59.                          missing.
  60.                   </li>
  61.                   <li >Bottle Retrieval Area is
  62.                          full.
  63.                   </li>
  64.                   <li >Bottle is dropped.
  65.                   </li>
  66.                   <li >Robot arm is not working.
  67.                   </li>
  68.                 </ul>
  69.                 <p>If an alarm and/or alert appears, refer to the instructions displayed
  70.                   on the screen to resolve the issue. If the problem persists, contact your local
  71.                   
  72.                   <ph
  73.                         conref="../../../../common/col_CompanyNames.xml#concept_A4D24345167D40F18313BEF25CD88553/ph_bioMerieux"><?xm-replace_text Phrase?></ph>
  74.                   Service Representative.
  75.                 </p>
  76.          </postreq>
  77.   </taskbody>
  78. </task>

advertising

Update the Post

Either update this post and resubmit it with changes, or make a new post.

You may also comment on this post.

update paste below
details of the post (optional)

Note: Only the paste content is required, though the following information can be useful to others.

Save name / title?

(space separated, optional)



Please note that information posted here will expire by default in one month. If you do not want it to expire, please set the expiry time above. If it is set to expire, web search engines will not be allowed to index it prior to it expiring. Items that are not marked to expire will be indexable by search engines. Be careful with your passwords. All illegal activities will be reported and any information will be handed over to the authorities, so be good.

worth-right
worth-right
fantasy-obligation